VALIDASI METODE PENENTUAN SAKARIN MENGGUNAKAN KROMATOGRAFI CAIR KINERJA TINGGI

Agus Kembaren, Tiasina Harahap

Abstract


Abstract. This research was aimed to validation analytical method of Saccarine and to determine saccarine consentration in soft drink High Performance Liquid Chromatography. Sodium Saccarine salts are used as standard.  Sodium Saccarine salts  dissolve in mobile phase methanol : fosfat buffer (60:40). Concentration of standard saccarine solution made in 20, 40, 60, 80, and 100 ppm. HPLC Agilent Technology 1260 Infinitely Series, Poroshell/C18 column-  50 mm lenght- id 4.6 mm - particles size 2.7 µm, UV-detector in λ 265 nm, injection volume 20 µL , and flow rate 1 mL/minutes were used in this research. Method validation parameters were determined by linierity, precision, accuracy, LOD, LOQ, and selectivity. The results show that lineariy with R = 0.998702, precision with RSD = 0.192%, accuracy with % recovery = 100.26%, LOD = 0.172 ppm, LOQ = 0.573, and selectivity with Rs = 0.2. Testing for the ‘Sprite Zero’ soft drink show that 117.6 mg/kg saccarine (under the maximum content in carbonated/noncarbonated soft drink). Conclution from this research is,  HPLC method can use in analysis saccarine content in soft drink

Kata kunci: metode penentuan,  sakarin, kromatografi kinerja tinggi


Full Text:

PDF


DOI: https://doi.org/10.24114/jpkim.v6i2.5590

Article Metrics

Abstract view : 2547 times
PDF - 2431 times

Refbacks

  • There are currently no refbacks.


Copyright (c) 1970 Agus Kembaren, Tiasina Harahap

Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 International License.

Jurnal Pendidikan Kimia
Contact: +62 853-1769-2813
Email: jpkim.pps@unimed.ac.id

Jl. Willem Iskandar, Pasar V, Medan Estate, Medan City, North Sumatra Province, Postal Code 20221. Phone/fax: (061) 661 3365 / +62 852-7802-1981.

Creative Commons License

This work is licensed under a Creative Commons Attribution 4.0 International License.